Turkish Journal of Chemistry

DOI

10.3906/kim-1109-42

Abstract

A new polyvinyl chloride (PVC) membrane electrode for determining the concentration of sodium dodecylsulfate (SDS) based on a neutral ion pair complex of dodecyltrimethylammonium-dodecylsulfate (DTA^+DS^-) as ionophore is described in this paper. The construction and characteristic performance of the PVC membrane electrode displays a Nernstian slope of --56.9 \pm 0.3 mV/decade in the concentration range of 2.0 \times 10^{-6} to 3.2 \times 10^{-3} mol L^{-1} and a detection limit of 1.6 \times 10^{-6} mol L^{-1}. The lifetime of the electrode is more than 3 months without showing significant change in the values of slope or linear concentration range. The electrode has wide pH range of application and short response time. The electrode shows selective response to DS^- and poor response to common anions except DBS^-. The electrode was used as an end point indicator electrode for the determination of SDS content in sodium dodecylsulfate/nonylphenol ethoxylates (SDS/NP-10) mixture solution by potentiometric titration using Hyamine standard solution as titrant. The maximum mol ratio (n_{(NP-10)}: n_{(SDS)}) when SDS cannot be determined in SDS/NP-10 mixture solution can be obtained.

Keywords

Anionic surfactant selective electrode, anionic surfactant, mixture solution, potentiometric titration

First Page

545

Last Page

555

Included in

Chemistry Commons

Share

COinS