•  
  •  
 

Turkish Journal of Medical Sciences

DOI

10.3906/sag-1909-58

Abstract

Background/aim: FloTrac/Vigileo is a noncalibrated arterial pressure waveform analysis for cardiac index (CI) monitoring. The aim of our study was to compare the CI measured by the 4th generation of FloTrac with PiCCO in septic shock patients. Materials and methods: We simultaneously measured the CI using FloTrac (CIv) and compared it with the CI derived from transpulmonary thermodilution (CItd) as well as the pulse contour-derived CI using PiCCO (CIp). Results: Thirty-one septic shock patients were included. The CIv correlated with CItd (r = 0.62, P < 0.0001). The Bland-Altman analysis showed a bias of 0.14, and the limits of agreement were ?1.62?1.91 L/min/m2 with a percentage error of 47.4%. However, the concordance rate between CIv and CItd was 93.6%. The comparison of CIv with CIp (n = 352 paired measurements) revealed a bias of -0.16, and the limits of agreement were ?1.45?1.79 L/min/m2 with a percentage error of 44.8%. The overall correlation coefficient between CIv and CIp was 0.63 (P < 0.0001), and the concordance rate was 85.4%. Conclusion: The 4th generation of FloTrac has not acceptable agreement to assess CI; however, it has the ability to tracked changes of CI, when compared with the transpulmonary thermodilution method by PiCCO.

Keywords

Cardiac output, hemodynamic monitoring, pulse contour analysis, arterial waveform analysis, measurement technique

First Page

860

Last Page

869

Share

COinS