Turkish Journal of Chemistry

DOI

10.3906/kim-2007-7

Abstract

In this study, quantative nuclear magnetic resonance (qNMR) method was used to determine the content of rosuvastatin in tablet. Linearity, range, limit of detection (LOD), limit of quantification (LOQ), accuracy, and precision were determined in validation study of rosuvastatin. Furthermore, validation study of rosuvastatin was performed with high performance liquid chromatography (HPLC). Uncertainties of qNMR and HPLC methods were determined using per EURACHEM/CITAC Guide CG 4 (3th edition), quantifying uncertainty in analytical measurement. qNMR and HPLC methods were linear in the ranges of 0.10 - 5.00 mg/mL and 0.001 - 0.0995 mg/mL, respectively and these lineraties indicate very good linearity performance with regression coefficients (R2 value) above > 0.99. Moreover, LOD and LOQ values using qNMR method were observed as 0.25 mg/mL and 0.80 mg/mL, respectively. These values using HPLC method were found as 0.00051 μg/mL and 0.001695 μg/mL, respectively. The strengths and weaknesses of qNMR method and HPLC method were determined with spectral emphasis on the role of identical reference standards in qualitive and quantitative analyses. It was found that qNMR method is simple, efficient, reliable, and accurate method. Moreover, qNMR method is an easy, practical, and useful method for the validation and optimization of rosuvastatin in the tablet.

Keywords

qNMR, proton nuclear magnetic resonance, high performance liquid chromatography, rosuvastatin, validation

First Page

132

Last Page

142

Included in

Chemistry Commons

Share

COinS