Turkish Journal of Chemistry
DOI
-
Abstract
All solid-state contact poly vinyl chloride (PVC) membrane potassium-selective electrode has been developed. The construction of the ion selective electrode membrane was organized using ionophore 4,4'-bis[4''-phenoxy(15-crown-5) methyl]benzyl in PVC cocktail. The potentiometric performance characteristics of the electrode were investigated by a computerized system. The electrode was found highly selective to K^+ ions over a wide variety of other cations. The electrode almost exhibited a Nernstian response (\ge 55 (\pm15) mV/decade of K activity) to K^+ within the concentration range 1.5 x 10^{-1} - 1.0 \times 10^{-5} M. The operational pH range of the electrode was 5--7. The response time of the electrode was fast (
Keywords
Ion-selective electrodes, Solid-state contact PVC-membrane, K^+ determination.
First Page
521
Last Page
528
Recommended Citation
KEMER, BARIŞ and ÖZDEMİR, MUSTAFA (2008) "Potentiometric Utility of the New Solid-State Sensor Based on Crowned Ionophore for the Determination of K^+," Turkish Journal of Chemistry: Vol. 32: No. 5, Article 1. Available at: https://journals.tubitak.gov.tr/chem/vol32/iss5/1