Turkish Journal of Chemistry

DOI

-

Abstract

An optical sensor for the direct measurement of silver was developed. The basic element of this optode is the immobilization of diphenylthiocarbazone on a triacetyl cellulose membrane. This optode has a linear range of up to 40 \mu g mL^{-1} of Ag^+ ions, with a limit of detection of 0.01 \mu g mL^{-1}. The response time of the optode was 4-7 min, depending on the concentration of Ag^+ ions.

Keywords

Optical sensor, Ag^+ ions, diphenylthiocarbazone

First Page

249

Last Page

257

Included in

Chemistry Commons

Share

COinS