Turkish Journal of Chemistry
DOI
-
Abstract
A biosensor for specific determination of hydrogen peroxide was developed using catalase (EC 1.11.1.6) in combination with a mercury film electrode. Catalase was immobilized with gelatine by means of glutaraldehyde on the electrode surface. The biosensor response was monitored by following the reduction peak of dissolved oxygen at around -0.24 V. The peak current changes upon addition of hydrogen peroxide and gives a linear response in a concentration range of 1-50 x 10^{-6} M within a response time of 3 minutes.
First Page
95
Last Page
100
Recommended Citation
ERTAŞ, F. NİL; TİMUR, SUNA; AKYILMAZ, EROL; and DİNÇKAYA, ERHAN (2000) "Specific Determination of Hydrogen Peroxide With A Catalase Biosensor Based on Mercury Thin Film Electrodes," Turkish Journal of Chemistry: Vol. 24: No. 1, Article 11. Available at: https://journals.tubitak.gov.tr/chem/vol24/iss1/11